<ErrorNumber>22107</ErrorNumber>
<ErrorFamily>Autoloader 2</ErrorFamily>
<ErrorMessage>
ALE2: Drum Vacuum Pressure to low to hold plate - please check vacuum system.
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>ThermoFlex</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
Check the sensor calibration:
Perform a drumvac cal and see if the problem goes away.
Check that internal sensing lines are connected and not leaking.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Verify that pump is on and hose is connected to machine.
Call for Service.
</text></DescriptionText>
</OperatorAction>
<Cause>
The level of suction in the drum is below the setpoint required to load the plate.
This can be caused by a number of things:
- Sensor mis-calibrated
- Sensing tube disconnected
- Vacuum pump hose disconnected
- Power lost at vacuum pump
</Cause>